<page xmlns="http://projectmallard.org/1.0/"
type="topic" style="task"
id="cursors">
<info>
<link type="guide" xref="index#analysis"/>
<link type="seealso" xref="measurements"/>
<desc>The shared vertical cursor pair and the per-row horizontal rulers, their four modes,
and how a reading is formatted.</desc>
<credit type="author copyright">
<name>Martino Ferrari</name>
</credit>
</info>
<title>Cursors</title>
<p>Cursors read values off the traces. There are two kinds: one <em>shared vertical
pair</em> that every row carries at the same time, and <em>horizontal rulers</em>, one pair
per row, that belong to that row alone.</p>
<section id="reach">
<title>How to reach it</title>
<steps>
<item>
<p>Press the cursors button in the header bar, or <key>c</key>. It switches the cursors on
and off.</p>
</item>
<item>
<p>The readings are in the dock (a read-only copy) and on the <gui>Cursors</gui> page, which
<gui>Edit</gui> beside <gui>Cursors</gui> opens over the plot and which owns the mode
and a <gui>Reset</gui> button. Drag a marker to move it.</p>
</item>
<item>
<p>The mode dropdown offers four modes: <gui>vertical</gui> (the shared time pair),
<gui>horizontal</gui> (the per-row rulers), <gui>both</gui>, and <gui>mix</gui>.</p>
</item>
</steps>
</section>
<section id="knowing">
<title>Worth knowing</title>
<list>
<item><p>The vertical pair is <em>one</em> time for the whole figure, so moving it on
any row moves it on all of them. That is what makes <code>dt</code> meaningful across
rows.</p></item>
<item><p>A ruler belongs to its row and nothing shares it, so two rows' rulers are
independent. The plots of a split row are one row here: they share their y axis, so the
rulers are at the same values in each.</p></item>
<item><p><gui>mix</gui> is the reading mode: one line per <em>bound</em> trace, with the
trace's own values at each marker and their difference. A slot with a name but no
channel behind it is not a trace and gets no line.</p></item>
<item><p>A value is the <em>nearest stored sample's</em>, never a number between two
samples. Zoom in past the sample spacing and the line between two samples is something
the plot drew, not data — so a reading off it would be the drawing. The hover crosshair
reads the same way.</p></item>
<item><p>Every reading carries a <em>metric prefix</em> so the digits sit in
<code>1..999</code>: <code>10.52 mV</code>, not <code>0.01 V</code>. Two decimals is all
an instrument landing on an <code>f32</code> sample can claim, so without the prefix a
small value would read <code>0.00</code> and the reading would be gone. A stream that
declares no unit gets the prefix anyway, written tight against the number
(<code>585.24m</code>) — the space separates the number from its unit, and a spaced
<code>m</code> with nothing after it would read as metres.</p></item>
<item><p>A marker past the newest sample reads <code>--</code>. With only one marker
over data the row reads <code>partial</code>: a difference between a number and nothing
is not a number.</p></item>
<item><p>In <gui>mix</gui> each marker also places a crossing annotation per trace,
which is what lets you see on the plot which sample a number came from. It is <em>on
that sample</em>, so when the marker falls between two it sits a little off its own
line: the line is where you put the marker, the dot is the datum being quoted. The hover
crosshair's mark snaps the same way, and is drawn larger than a trace's sample dots so
it does not read as one more sample.</p></item>
<item><p>The time row exists only in the modes that show the vertical pair.
<gui>horizontal</gui> has no time row, by design.</p></item>
<item><p>A cursor <em>stays where you put it</em> while a live view rolls under it,
which is what a bench scope does: its <code>T0</code> is an age, so a marker left at
<code>-1.5s</code> keeps reading the sample a second and a half old as the stream
advances. A measurement's marks do the opposite — they travel with the sample they
measured — because a cursor is a question you asked and a mark is an answer about one
particular sample.</p></item>
</list>
</section>
<section id="example">
<title>Example</title>
<p>Three traces in <gui>mix</gui> mode with both markers over data: 3 lines and 6 crossing
marks, each line reading like <code>sine 0.31 V -0.84 V -1.15 V</code>. Move the second
marker past the newest sample and the third column reads <code>--</code>.</p>
</section>
</page>